• ISSN 0258-2724
  • CN 51-1277/U
  • EI Compendex
  • Scopus
  • Indexed by Core Journals of China, Chinese S&T Journal Citation Reports
  • Chinese S&T Journal Citation Reports
  • Chinese Science Citation Database
Volume 27 Issue 3
May  2014
Turn off MathJax
Article Contents
YANG Jie, YANG Ji, XU Jiang, LIU Yujie, HUANG Nan. Verification of Stent Mechanical Formulas by “Planar Stent” Experiment[J]. Journal of Southwest Jiaotong University, 2014, 27(3): 519-524. doi: 10.3969/j.issn.0258-2724.2014.03.023
Citation: YANG Jie, YANG Ji, XU Jiang, LIU Yujie, HUANG Nan. Verification of Stent Mechanical Formulas by “Planar Stent” Experiment[J]. Journal of Southwest Jiaotong University, 2014, 27(3): 519-524. doi: 10.3969/j.issn.0258-2724.2014.03.023

Verification of Stent Mechanical Formulas by “Planar Stent” Experiment

doi: 10.3969/j.issn.0258-2724.2014.03.023
  • Received Date: 08 Jan 2013
  • Publish Date: 25 Jun 2014
  • To verify the formulas of elastic radial stiffness and plastic limit pressure of stents, the conception of "planar stent cell" was proposed and a planar stent specimen was produced according to the real curved structure of an intravascular stent. Tensile experiments of "planar stent cell" were then performed using MTS (material test system) equipment, and the experimental pressure-radial curves were compared with the ones predicted by formulas of elastic radial stiffness and plastic limit pressure. The results show that at the elastic stage of stent expansion, the difference of elastic radial stiffness between experimental measurements and formula-calculated values was less than 3%; at the plastic stage, the two curves obtained by experiment and formula were close. The experiments confirmed the accuracy and precision of the two formulas.

     

  • loading
  • ETAVE F, FINET G, BOIVIN M, et al. Mechanical property of coronary stents determined by using finite element analysis[J]. J. Biomechanics, 2001, 34: 1065-1075.
    CHUA S N D, MACDONALD B J, HASHMI M S J. Effects of varying slotted tube (stent) geometry on its expansion behaviour using finite element method[J]. J. Materials Processing Technology, 2004, 155-156: 1764-1771.
    MIGLIAVACCA F, PETRINI L, MONTANARI V, et al. A predictive study of the mechanical behaviour of coronary stents by computer modeling[J]. Med. Eng. Phys., 2005, 27: 13-18.
    WALKE W, PASZENDA Z, FILIPIAK J. Experimental and numerical biomechanical analysis of vascular stent[J]. J. Materials Processing Technology, 2005, 164-165: 1263-1268.
    GERVASO F, CAPELLI C, PETRINI L, et al. On the effects of different strategies in modeling balloon-expandable stenting by means of finite element method[J]. J. Biomechanics, 2008, 41: 1206-1212.
    GARC A A, PE A E, MART NEZ M A. Influence of geometrical parameters on radial force during self-expanding stent deployment. application for a variable radial stiffness stent[J]. Journal of the Mechanical Behavior of Biomedical Materials, 2012, 10: 166-175.
    王晓, 冯海全, 王文雯, 等. 球囊扩张式冠脉支架生物力学性能研究[J]. 中国生物医学工程学报, 2013, 32: 203-210. WANG Xiao, FENG Haiquan, WANG Wenwen, et al. Research on biomechanics properties for balloon-expandable intracoronary stents[J]. Chinese Journal of Biomedical Engineering, 2013, 32: 203-210.
    WANG R, CHANDAR K R. Mechanical response of a metallic aortic stent-part Ⅰ: pressure-diameter relationshiop[J]. Journal of Applied Mechanics, 2004, 71: 697-705.
    程洁, 周啸, 李俐军, 等. 冠脉支架的多功能体外力学性能测试装置及实验研究[J]. 东南大学学报:自然科学版, 2010, 40(2): 341-345. CHENG Jie, ZHOU Xiao, LI Lijun, et al. In-vitro test apparatus and experimental study of mechanics properties of coronary stents[J]. Journal of Southeast University: Natural Science Edition, 2010, 40(2): 341-345.
    YANG J, HUANG N. Mechanical formula for the plastic limit pressure of stent during expansion[J]. Acta Mechanica Sinica, 2009, 6: 795-801.
    YANG J, HUANG N. Formula for elastic radial stiffness of the tubular vascular stent[C]//6th World Congress on Biomechanics. Singapore:[s.n.], 2010: 1435-1438.
    DOMOULIN C, COCHELIN B. Mechanical behaviour modelling of balloon-expandable stents[J]. Journal of Biomechanics, 2000, 33: 1461-1470.
    PARK W P, CHO S K, KO J Y, et al. Evaluation of stent performances using FEA considering a realistic balloon expansion[J]. Proceedings of World Academy of Science: Engineering & Technolog, 2008, 37: 117-122.
    DUERIG TW, WHOLEY M. A comparison of balloon-and self-expanding stents[J]. Min Invas Ther & Allied Technol, 2002, 11(4): 173-178.
    MAEKAWA Y, ASAKURA Y, ANZAI T, et al. Relation of stent overexpansion to the angiographic no-reflow phenomenon in intravascular ultrasound-guided stent implantation for acute myocardial infarction[J]. Heart Vessels, 2005, 20: 13-18.
    ALEXANDRE Q, CARLOS A M G, EMERSON C P. Hyperexpansion of coronary stents and clinical outcomes[J]. Texas Heart Institute Journal, 2006, 33: 437-444.
    GREGG W S, FREDERICK G G, JOHN M H. Analysis of the relation between stent implantation pressure and expansion[J]. The American Journal of Cardiology, 1999, 83: 1397-1400.
  • 加载中

Catalog

    通讯作者: 陈斌, bchen63@163.com
    • 1. 

      沈阳化工大学材料科学与工程学院 沈阳 110142

    1. 本站搜索
    2. 百度学术搜索
    3. 万方数据库搜索
    4. CNKI搜索
    Article views(749) PDF downloads(383) Cited by()
    Proportional views
    Related

    /

    DownLoad:  Full-Size Img  PowerPoint
    Return